Skip to content

OD600 Measurement Results Using Spectrophotometers & Online OD Instruments

Spectrophotometers can be an excellent tools to monitor the growth phase or determine cell concentrations in microbial fermentation processes. Despite of the fact that these instruments are generally utilized to measure absorbance of particular samples (concentration measurements), spectrophotometers are also used to quantify light scattering running at 600 nm wavelength. This method, commonly known as OD600 measurement, is known to provide consistent cell growth results when well-accepted practices are applied.

In addition, new online OD probes can be utilized to monitor a process continuously, using the OD600 spectrophotometer data to validate the online results and calibrate the online measurements in concentration units (i.e. grams / liter).

An OD600 measurement is dependent on the type of instrument configuration being used, and also differs on its results comparing among different microbial processes (organisms). For example, larger organisms tend to scatter more light than smaller ones (yeast vs bacteria), and minor differences in cuvettes’ paths produce different measurement results. The utilization of the proper measurement technique allows the user to produce results that are comparable among different instrument configurations and that can be trusted as a repeatable method to qualify a particular process.

Table of Contents

OD600 Measurement Principle

As visible (600 nm) light passes through the sample, it is scattered in a 360 degree radius with a Radom distribution of intensities depending on cell shape and size and on the concentration of cells in the sample. A measurement of the light intensity reflected by the sample will then correlate to the cell concentration. The path length and wave length on the particular instrument setup must be well-defined and controlled to secure reproducible results. Additionally, since researchers rely on OD600 values published for particular processes to establish the right time for harvesting or inoculation, it is essential to pay attention to the correct instrument parameters.

UV Vis spectrophotometer
Example of a UV/Vis spectrophotometer that provides the ability to measure samples in standard cuvettes and / or micro samples

Available through PROAnalytics: UV VIS Spectrophotometer

Empirical Values

Scientists are often relying on published OD600 data to take different actions during the course of a fermentation. For optimal results, the OD600 measurements need to be consistent and repeatable.

Since all the variables related with the instruments being used and the processes being characterized are rarely known, it is recommended that empirical concentration measurements be determined by the use of growth curves correlating OD600 values against plate counts for any particular cell type.

New UV/VIS spectrophotometers provide the ability of measuring micro samples extending the measurement range to high concentration values. This is true thanks to the automatic path length adjustment feature. It is, however, recommended to establish a conversion factor comparing results among spectrophotometers, especially in cases where historical data has been generated with cuvette-type instruments.

The best approach for determining the conversion factor involves utilizing samples in both spectrophotometers that are closed to a desired “target” OD600 number specially relevant to the process of interest. In those cases where the cuvette-style instrument cannot reach a high concentration measurement, apply the correct dilution technique. Once both measurements have been obtained, the conversion factor is calculated as the ratio between both measurements:

OD600 spectrophotometer conversion factor

Comparative Study

A widely accepted Turbidity Standard (1000 NTU) can be used to compare results among two types of spectrophotometers. As mentioned earlier, dissimilar optical arrangements among two instruments would provide different OD results when measuring the same sample for scattered light. To test and recognize the differences, and ultimately establish a conversion factor, the 1000 NTU standard Formazin solution can be subsequently diluted by a factor of 2, registering each measurement (for each dilution) on both instruments.

After performing 7 dilutions and obtaining a total of 7 measurements on each instrument, the conversion factor can be calculated as the ratio between the average reading on Spec A to the average reading on Spec B. Furthermore, the obtained results can be utilized to developed the linear relationships between OD600 and NTU for each spectrophotometer.

Additionally, if the ultimate objective is to establish a correlation between an OD600 spectrophotometer measurement and an Online Optical Density Probe, each sample measurement on the OD probe can be recorded as well to also establish that comparison.

Table 1: to be filled in with experimental results

Establishing a correlation between OD probe and spectrophotometer results
OD600 measurement instrument BugLab BE3000
Example of an online Optical Density probe designed to operate continuously under standard process conditions

Available through PROAnalytics: OD Measurement Instruments

Linear Range Measurements

Spectrophotometers using cuvettes are limited in the upper concentration range. Generally the highest measurable OD600 value to achieve with these types of instruments maintaining linearity is not very high.

To compare readings against micro-sample spectrophotometers (which reach much higher concentration values), it is necessary to dilute the samples to be measured in the cuvette-type instrument. At the same time, if correlating with live cell concentration measurements (like those from plate counts ), running samples from different stages of the growth curve on the cuvette spectrophotometer will show the maximum concentration value that can be measured before losing linearity in the measurement.

Once all data has been collected for each particular sample concentration, the two different linear relationships (for Spec A and Spec B) can be plotted. Slopes will generally differ. Experimentally, applying good sampling and analytical techniques, the variation between the two slopes is in the order of 15- 25%. At the same time, the Online OD probe results can be plotted to observe how well the discrete spectrophotometer results fit this third linear relation. The spectrophotometer results can be used as well to create a calibration curve on the Online OD probe instrument.

Determination of Concentration Values

As mentioned earlier, different cell types create distinct light scattering profiles resulting in different correlation factors between OD600 values and viable cell concentration values. For that reason, it is essential to establish a dedicated correlation factor for each particular process/cell type. Modern spectrophotometers and OD600 measurement instruments allow a user to enter the calibration results to display concentration values directly.

Generally accepted practices for OD600 spectrophotometer measurements:

  • When taking a sample from your culture, make sure the process is well mixed and that no cell settling is taking place.
  • Properly cleaned high quality cuvettes should be used to avoid errors associated to the analytical technique. Mettler Toledo recommends using Quartz Cuvettes only.
  • Do not use glass pipettes. They can scratch the quartz cuvettes.
  • Insert the cuvettes with the proper orientation in the instrument.

When using a micro sample spectrophotometer:

  • Thoroughly clean both sample measurement surfaces before introducing the sample.
  • Pipette 1 micro liter of sample making sure no air bubbles are introduced.
  • Close the arm sealing the chamber tightly.

In summary, the above technical note is intended to provide a basic method for obtaining repeatable OD600 measurement results on particular fermentation processes and correlate those values with online OD probes measurements with the ultimate objective of establishing a reliable process calibration curve.

This entry was posted in Reports. Bookmark the permalink.

玻璃钢生产厂家玻璃钢红军雕塑定制蚌埠定制玻璃钢雕塑销售厂家新乡哈萨克玻璃钢人物雕塑临沧市玻璃钢雕塑设计哪里买玻璃钢雕塑加工有废水鞍山出口沈阳玻璃钢花盆山西玻璃钢抽象景观雕塑商场装饰玻璃钢人物雕塑货源充足汕尾玻璃钢人物雕塑玻璃钢雕塑海豚商场秋天美陈照相背景图片徐州玻璃钢雕塑定制美陈玻璃钢动物雕塑销售厂家潮州玻璃钢雕塑推荐货源合肥玻璃钢雕塑优点泉州玻璃钢花盆厂固原玻璃钢雕塑加工广州特色玻璃钢雕塑价位长沙标识玻璃钢彩绘雕塑厂家安庆水果玻璃钢雕塑商场枫叶美陈讲解佛像玻璃钢雕塑价位山东玻璃钢企鹅雕塑大同玻璃钢人物雕塑定制价格湖南怀化玻璃钢雕塑制作运城玻璃钢花盆广州人物玻璃钢雕塑定制玻璃钢雕塑开裂是什么原因白城酒店玻璃钢雕塑工厂江苏个性化玻璃钢雕塑图片香港通过《维护国家安全条例》两大学生合买彩票中奖一人不认账让美丽中国“从细节出发”19岁小伙救下5人后溺亡 多方发声单亲妈妈陷入热恋 14岁儿子报警汪小菲曝离婚始末遭遇山火的松茸之乡雅江山火三名扑火人员牺牲系谣言何赛飞追着代拍打萧美琴窜访捷克 外交部回应卫健委通报少年有偿捐血浆16次猝死手机成瘾是影响睡眠质量重要因素高校汽车撞人致3死16伤 司机系学生315晚会后胖东来又人满为患了小米汽车超级工厂正式揭幕中国拥有亿元资产的家庭达13.3万户周杰伦一审败诉网易男孩8年未见母亲被告知被遗忘许家印被限制高消费饲养员用铁锨驱打大熊猫被辞退男子被猫抓伤后确诊“猫抓病”特朗普无法缴纳4.54亿美元罚金倪萍分享减重40斤方法联合利华开始重组张家界的山上“长”满了韩国人?张立群任西安交通大学校长杨倩无缘巴黎奥运“重生之我在北大当嫡校长”黑马情侣提车了专访95后高颜值猪保姆考生莫言也上北大硕士复试名单了网友洛杉矶偶遇贾玲专家建议不必谈骨泥色变沉迷短剧的人就像掉进了杀猪盘奥巴马现身唐宁街 黑色着装引猜测七年后宇文玥被薅头发捞上岸事业单位女子向同事水杯投不明物质凯特王妃现身!外出购物视频曝光河南驻马店通报西平中学跳楼事件王树国卸任西安交大校长 师生送别恒大被罚41.75亿到底怎么缴男子被流浪猫绊倒 投喂者赔24万房客欠租失踪 房东直发愁西双版纳热带植物园回应蜉蝣大爆发钱人豪晒法院裁定实锤抄袭外国人感慨凌晨的中国很安全胖东来员工每周单休无小长假白宫:哈马斯三号人物被杀测试车高速逃费 小米:已补缴老人退休金被冒领16年 金额超20万